Zeiss LSM 700 Instructions Manual
Hide thumbs Also See for LSM 700:

Advertisement

 
Table of contents 
 
Introduction ...................................................................................................................................... 2
Hardware .......................................................................................................................................... 2
Objectives ...................................................................................................................................... 2
Filter turret .................................................................................................................................... 2
Laser lines ...................................................................................................................................... 3
Light Path ...................................................................................................................................... 3
Start up ............................................................................................................................................. 4
Sample mounting and viewing ......................................................................................................... 5
Software ........................................................................................................................................... 5
Sample viewing .............................................................................................................................. 6
Acquisition ..................................................................................................................................... 8
LSM Channel mode confocal setting ......................................................................................... 8
Configuration: ........................................................................................................................ 8
Single image acquisition ........................................................................................................ 12
Multidimensional acquisition ................................................................................................. 12
Shutting down ................................................................................................................................. 17
 
 
 
 
 
Instruction LSM700 2017-12-07.docx   2017-12-19
Confocal LSM 700
 
 
 
 

Advertisement

Table of Contents
loading

Summary of Contents for Zeiss LSM 700

  • Page 1: Table Of Contents

        Instruction LSM700 2017‐12‐07.docx   2017‐12‐19 Confocal LSM 700     Table of contents    Introduction ............................2 Hardware ............................2 Objectives ............................2 Filter turret ............................ 2 Laser lines ............................3 Light Path ............................3 Start up ............................. 4 Sample mounting and viewing ......................5 Software ............................5 Sample viewing ..........................6 Acquisition ............................. 8 LSM Channel mode confocal setting ..................
  • Page 2: Introduction

    Introduction  The Zeiss LSM 700 is equipped with 4 laser lines and can detect up to four color signals at frame rates  approaching 5 frames per second at 512 x 512 pixels. Efficient separation of the fluorescence signals by  selective laser excitation coupled to efficient splitting of the emission using the variable secondary  dichroic (VSD) beamsplitter prevents crosstalk and enables spectral imaging as well as linear unmixing of  highly overlapping fluorophores. Among the advanced features of the VSD beamsplitter is that all portions  of the emission spectrum are utilized for determining each spectral data point.  Hardware  The microscope:  The LSM 700 is attached to an upright microscope (Axio Imager Z2).  Objectives    Working  Parfocal  Position  Objective lens Magnification NA coverslip Immersion  distance  length  1  W Plan  45 mm  X40  0  2.5 mm  Water dipping  Apochromat  2  EC Plan  45 mm  0.3 0.17 mm 5.2 mm  Air  Neofluar  3  Plan ...
  • Page 3: Laser Lines

    Laser lines    Laser  Line Solid state 5mW   639  Solid state 10mW  555  Solid state 10mW  488  Solid state 5mW  405    Light Path      1. The lasers come aligned from the manufacturer, all 4 lasers are attached to an optic cable and  enter the confocal scan head. (1)  2. The laser lines are reflected by the main dichroic mirror (beamsplitter). (4).  3. The 2 Galvanometric mirrors scanthe sample. (3).  4. The return emission wavelength passes through the main dichroic mirror (4).   5. The emission passes through the pinhole aperture and the out of focus light is eliminated.(5)  6. The variable secondary dichroic beamsplitter (VSD) is coated with a variable coating that between  420nm ‐ 630nm, according to the position of light incidence determines the cutoff below which   wavelength the light will pass to PMT1 and above this wavelength it will be reflected towards  PMT2.(6)  7. Before the detection an emission filter can be inserted from the filter wheel before the PMT, only  longpass (LP)/shortpass (SP) filters are available. (9)  8. The light is detected in one of the two PMT detectors‐ PMT 1 for SP emission wavelength, PMT 1  for LP emission wavelength.(7,8)      p. 3 ...
  • Page 4: Start Up

    Start up  Turn on the system in the following order:  ‐ In the power strips turn on switches 1  2.    ‐ Turn on the stage controller (3)    ‐ Rotate (clockwise) the silver key in the lasers electronic unit(4)    ‐ Turn on the computer and log in  User name: multilabs  Password: 123456  ‐ Turn on the metalhalide lamp (6)  6          p. 4   ...
  • Page 5: Sample Mounting And Viewing

    Sample mounting and viewing  Insert the slide or dish to the stage holder.  Push in the Light switch between the ocular and the confocal.     Via the touch pad:  1. Choose the preferred objective       2. Choose the proper filter set and open the TL\RL shutter:    Filter set    Software  ‐   Log into BookitLab and activate your reservation to start Black Zen 2.3  ‐   To acquire images click Start System       p. 5   ...
  • Page 6: Sample Viewing

    Sample viewing   In the Locate tab choose the preferred objective filter set and open the illumination shutter  RL\TL        Set the Transmitted light (In case the TL is needed)  Click the transmitted light lamp icon, click ON and set desired brightness (or via the  microscope).  Set the filter turret position to DIC and the condenser below the stage to I, II or III according  to the objective lens (see objective lens info on touchpad).  If you are acquiring transmitted channel set up Kohler illumination:    Bring the sample into focus. Close the field stop using the motorized aperture button (F) located at the right of the    microscope until you can see at least one edge   Adjust the condenser height until the edges of the diaphragm image are crisp.   Center the diaphragm image using the two centering screws.   Open the field diaphragm, just until the image fills the field of view.    p. 6   ...
  • Page 7          2, 5 3    p. 7   ...
  • Page 8: Acquisition

    Acquisition  LSM Channel mode confocal setting  In the software go to the Acquisition main tool tab     Pull out the Light switch between the ocular and the confocal.     Configuration:  It is highly recommended to seek BCF staff’s advice for new fluorophores (exact name) or new  combinations.    Choose one:   Load a saved configuration from the Experiment Manager          Reuse settings of an image: open an image and below the image click Reuse      New setup: click  , choose the desired dyes from the list, choose the scanning  method and click Apply                               ...
  • Page 9  Select show all tool   and open the Imaging set up tab  Note: make sure that the chosen configuration is correct and suits the fluorophores; make  changes if needed it is highly recommended to check all dyes spectrums in a spectra  viewer  ThermoFisher  Chroma  6  3.1    1. Laser lines  2. Main dichroic mirror  3. PMTs: PMT1 and PMT2  4. Variable secondary dichroic beamsplitter (VSD).  5. Emission Filter selection   6. Emission spectrum with Graphic representation of the laser line, the position of the  VSD and emission filter.  7. Transmitted light PMT‐ a transmitted light image (non confocal) can be added to the  confocal image (it recommended to add to the longest wavelength track).        p. 9   ...
  • Page 10: Acquisition & Parameter Setting

    Acquisition & parameter setting  Pull out the lightpath switch between the ocular and the confocal   Parameter setting   Open the Acquisition mode tab     Scan mode:  frame (in most cases);                    line scan mode can be useful for physiology experiments.   Frame size:   512X512   Speed:    8    Averaging:     Method:  mean (in most cases):   in low signal with low noise Sum method may be useful   Number:  1  Mode:  frame for fixative sample, for live cells and in case of unstable dye line  mode will be more suitable  Bit Depth:  defined the dynamic range of the image, choose 8bit\12 bit according to  the biological question.   Direction:  ‐‐‐‐> (more precise)   Scan area:  slight nudge, zoom and rotation      ...
  • Page 11  Open the Channels tab     Check and select only the shortest wavelength track and click the 1AU button    Start scan in continuous mode     Check the Range indicator   (below the image container in the Dimensions  tab).    Set your image parameters to achieve image without saturation (saturation=red pixels in  the image) and the correct black level (black= blue pixels in the image)  o Set the PMT gain to 550‐600.  o Set the Laser % ‐ it is recommended to use the lowest % as possible and to  increase gain (master).  o In case of low signal with high laser % and high gain you can add digital gain  (up to 2).  o Set the offset level until you reach the PMT threshold (blue pixel will appear in  the image while scanning in range indicator mode).   Stop the scan     Select another track; set its pinhole to the same μm section as the first track and set its  parameters and so on    Stop the scan          p. 11   ...
  • Page 12: Single Image Acquisition

    Single image acquisition   In the Acquisition Mode tab select the preferred zoom (alternatively you can easily  position the cropping frame to zoom in via the crop tool   below the image in the  Dimensions tab)     Scan one frame to check position.   Check all tracks in the Channel tab, In the Acquisition Mode tab click   to set the  frame size.    Set scanning speed to 6   Set averaging number to 2‐4 (according to the SNR)   Press Snap    in the start buttons    Save the image: press   in the Images and Documents tab (right hand side of the  window) and save in the directory:   D:\Users data\PI name\MM‐YYYY\User name\YYYY‐MM‐DD    Multidimensional acquisition     Z stack   In the Multidimensional Acquisition options below the start buttons check Z stack; a new  tab (Z‐stack) will appear under the Multidimensional Acquisition Section.   Set the interval: press Smallest for the system recommendation (Nyquist criterion)   With one channel checked and In the Acquisition mode tab:  Frame size    512X512  Scanning speed  8  ...
  • Page 13  Start scan in continuous mode    Define the signal boundaries by moving the focus wheels and pressing Set First in one  boundary and Set Last in the second boundary.     With all channels checked, in the Acquisition mode tab:   Frame size    Press optimal  Scanning speed  7  Averaging number  2‐4   Press      Save the Image  Note: there are more advanced Z stack options in this tab such as brightness correction  over z.  Time series   In the Multidimensional Acquisition options check Time series; a new tab (Time Series) will  appear under the Multidimensional Acquisition Section.   Define number of cycles and the interval  Note: in case of live cell imaging avoid high laser intensity      In the Acquisition Mode tab:   Frame size    Press optimal  In case of big frame size with long frame scan time it recommended to increase the  zoom and set a new frame size with less pixels by pressing optimal again.  Scanning speed:   7‐8  Averaging number  1‐2  ...
  • Page 14   Tile   In the Multidimensional Acquisition options check Tile scan; a new tab (Tile Scan) will  appear under the Multidimensional Acquisition Section.   Define the number of tiles   Define the overlap % ‐ recommended is 8%‐15%   Calculate slide rotation: In the main menu choose Macro  Tile scan rotation  In the popup window press Calibrate and wait until the process is finished (the rotation  angle will be exported to the tile scan tab)  then press close.      Note: Online stitching is not a recommended option, the process can be configured in the  processing main tool tab.   In the Acquisition mode tab:   Frame size:     Press optimal  Zoom:    It is recommended to set the zoom above 1.2  Scanning speed:  7  Averaging number:  2   Press      Save the Image   Stitching: In the Processing main tool tab    Stitch  select your input image   press Apply.  In case of uneven illumination the stitching process can be done with more options in  the ZEN blue software.   ...
  • Page 15 Repeat bleach after number of scans  The amount of iterations  The laser and its intensity  More setting can be defined in the bleaching dialog.      Define one or more ROIs:  After a single scan in the Regions tab choose one of the ROI tools and draw on the  acquired image; the dimensions can also be set manually in the Width/Height textboxes.   Define which process will be performed in the ROI – Acquisition/Bleach/Analysis.       Define Time series:    In the Acquisition mode tab:  Frame size    Press optimal  Scanning speed  7 or more  Averaging number  1‐2   Press       p. 15   ...
  • Page 16  Save the image   The intensity data from the ROI can be viewed in the ROI mean tab in the image window  Note: The region can be defined in other types of experiments with or without time lapse  and can also be used as a tool of zooming in by checking Fit Frame size to bounding  rectangle of regions.  Positions   A positions list can be added to a multidimensional experiment in which all of the settings  will be acquired (channel, acquisition mode, Z stack, tile, time series).   The Positions tab can be a useful tool in managing a large tile and for quick slide scanning  (via the ocular) with low magnification before moving to high magnification with small field  of view.     Add a position by pressing Positions below the image container and clicking on the image  container that was just acquired at the current stage position    Check Positions at the Multidimensional Acquisition options.   At the Positions tab move to a position in the list by selecting and pressing Move to   A position can also be added by pressing Add in the Positions tab   Note: If you use the positions list only to locate saved positions uncheck Positions from the  Multidimensional Acquisition options before pressing Start Experiment, otherwise your  settings will be acquired at every position in the list.     Note you can use multiple options in the multidimensional scanning for example Z stack & tile  Time series, bleaching, ROI & Z stack and positions.      p. 16   ...
  • Page 17: Shutting Down

    Shutting down   ‐ Make sure all the data is saved. The data will be automatically saved to the server during the  night. PLEASE DO NOT USE ANY USB FLASH DRIVE on this computer.  ‐ Clean any oil immersion objectives you used with lens paper and petroleum ether, twice.    ‐ Close the ZEN software  ‐ Log off from your account in the BookItlab window  ‐ Check if another user has a reservation; if yes verify his/her arrival. If you are the last user for the  day or the next reservation is in two hours or later, continue with system shutdown.  ‐ Turn off the metalhalide lamp (6)  ‐ Turn off the computer (5) only on weekends  ‐ Turn off the laser electronic unit (4)  ‐ Turn off the stage controller (3)  ‐ Turn off switches 2  1 In the switch box   ‐ Cover the microscope    p. 17   ...

Table of Contents