Zeiss LSM 880 Operating Manual page 583

Table of Contents

Advertisement

LSM 880
• The movement of the stage is defined
highlighting the Move stage action block,
chosing the Mode Relative Position in the
Property list and typing in the value 10 for X
and Y which then will move the stage 10
microns into X and Y direction starting with the
current stage position on passing that action
block.
• Now the first complete scan has been
performed and the image is stored. For the next
cycle the variable Index has to get another value
to make sure the numbering of the images
works. Using the Assignments action block,
the Variable Index is chosen and the value
assigned is Index + 1. Such the variable will get
the transient value 2 for the next cycle, the
value 3 for the over next cycle and so on.
• The number of repeats can be defined using a
Decision block and using the value of the
variable as base for the decision. By defining
that if the Expression of the variable is larger
than 5 and not connecting the output arrow
marked with "yes" to any other block, the
macro will stop after the acquisition of five
images. Alternatively one can use at this part of
the program flow the Repeat block and define
the number of repeats to be performed, which
would be 5 in this case (The first scan is already
counted!).
10/2014 V_01
CHAPTER 2 - MACROS AND VISUAL BASIC
Visual Macro Editor
000000-2071-464
Fig. 17
Property list of the Move stage action
block
Fig. 18
Property list of the Assignments action
block
Fig. 19
Property list of the Decision action
block
ZEISS
25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lsm 880 nlo

Table of Contents