− Account for overlap (multi-emitter): multi-
emitter events will be fitted taking into
account
the
Maximum cluster size slider and input box
become available (Fig. 443). Here you can
select the maximum number of molecules in
the cluster that are used for computation. So
a number of 10 means, for example, that
clusters with 10 or less peaks will be taken,
and thoise exceeding 10 will be discarded.
We recommend to limit custer size to 10 as
otherwise computation can become time
consuming.
White circles indicate that the particular emitters
are used (if channel color is grey; otherwise the
channel color is used), red circles that the emitters
are discarded for the fitting process. For account of
overlap the color stays white or in the channel
color regardless if a cluster may be discarded due
to the Maximum cluster size setting.
The Dimension drop down menu will display the
dimensionality (2D or 3D) of the selected PALM
image, which is automatically recognized if a .czi
format is loaded (Fig. 444). It is possible to overrule
the automatically chosen value.
In the Channel drop down menu you can select
the channel for which the Processing should be
conducted for a multi-channel image.
Expand the Peak Finder field to have access to the
peak selection filters (Fig. 445). Two filters are
available:
− Peak mask size
− Peak intensity to noise
Use the sliders or input spin boxes to set the
corresponding values.
We recommend to start with the default
values for the peak mask size (9 for 2D; 19
for 3D data) and then optimize on behalf of
the results. Note that in 3D-PALM using
PRLIM (phase ramp localization imaging
microscopy) with a double phase ramp, the
PSF is split into two lobes (
a larger expansion than the classical PSF.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
overlap.
If
selected,
) that have
Fig. 446
000000-2071-464
the
ig. 443
Maximum cluster size field
F
Fig. 444
Dimension drop down menu
Fig. 445
Peak Finder Expansion field
Fig. 446
Double lobed PSF created
by a double phase-ramp in
the pupil plane
ZEISS
331