Center Screen Area / Image Containers - Display and Image Analysis
When Immobile structures (+Stack) is selected, the average image from the stack (pixel by pixel) is
subtracted from every image and a scalar is added to avoid negative pixels. The scalar is computed from
the average of all pixels from all frames of the stack. When Immobile structures (+Plane) is selected,
the average image from the stack (pixel by pixel) is subtracted from every image and a scalar is added to
avoid negative pixels. The scalar is computed from the frame to which it is added. Both procedures are
equally valid in most of applications.
6.19.4

Display Functionalities

Activate Display under the RICS view. The Display options will be displayed in the View control window
(Fig. 713).
Fig. 713
Display control window
In the Representation pull down menu you can
choose how the correlation function should be
displayed.
Four options are available (Fig. 714):
Diagram will display the horizontal and vertical
profiles of the correlation function and the fit /
residuals if activated;
Image will display the correlation and the fit /
residuals if activated as two dimensional images;
Table will display the coordinates of the horizontal
and vertical profiles of the correlation and the fit /
residuals if activated that can be copied or saved
via a right mouse click from the Table Export
menu (Fig. 715);
2.5 D will display the correlation function and the
fit / residuals if activated as pseudo 3D images
(Fig. 716).
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Fig. 714
Fig. 715
000000-2071-464
Representation pull down menu
Table export menu
ZEISS
513