Zeiss LSM 880 Operating Manual page 616

Table of Contents

Advertisement

ZEISS
Connectors:
Pixelclock = Coax:
Pin
Signal name
Signal
PCLK_Out
Signal shield
GND
Outer shield
Shield
The pixel clock is exactly symmetrical to the pixel. For example if the pixel rate is 10 kHz the high-level
and the low level both have a time of 50 microseconds. The width of the TTL pulse therefore is 1/ (2 x
pixel rate).
15-point high density D-Type plug: Plug "A"
Pin
Signal name
1
SyncOUT0 synch output
2
SyncOUT1
3
SyncOUT2
4
SyncOUT3
5
GND
6
AsyncOUT0 asynch output
7
AsyncOUT1
8
AsyncOUT2
9
+5 V
10
GND
11
AsyncIN0 input
12
AsyncIN1
13
AsyncIN2
14
AsyncIN3
15
Connector detector*
14
CHAPTER 4 - ANNEX
Specification of Trigger-Interface
Triax Lemo S. A. Serie 00, EPL.00.650NLN
000000-2071-464
Description
PixelClock Output
Digital ground
Chassis ground
Description
Synchronous output
Synchronous output
Synchronous output
Synchronous output
Digital ground
Asynchronous output
Asynchronous output
Asynchronous output
Va
Digital ground
input
input
input
input
Input / Connector detector -
LSM 880
Software interface
Trigger 1 out
Trigger 2 out
Trigger 3 out
Trigger 4 out
-
-
-
-
-
-
Trigger 1 in
Trigger 2 in
Trigger 3 in
Trigger 4 in
10/2014 V_01