Zeiss LSM 880 Operating Manual page 633

Table of Contents

Advertisement

LSM 880
Motorized Periscope Panel
This panel provides the interface to control the motorized mirrors in the periscope
Beam 1 and Beam 2 represent the two input beams according to the following constellation:
Laser system
NLO laser + OPO
NLO laser + NLO laser
Input and Output correspond to the position of the mirror inside the periscope as described above.
Each axis of each mirror can be controlled independently. For each axis four buttons are available, two for
each direction. Axis 1 represents the vertical movement of the beam in the two directions. Axis 2
represents the horizontal movement of the beam in the two directions.
Buttons with one arrow cause a small step movement of the mirror along the corresponding axis.
Buttons with two arrows cause a large step movement of the mirror along the corresponding axis.
10/2014 V_01
CHAPTER 5 - MULTIPHOTON LASER SCANNING ...
NLO (Non Linear Optics) Laser for LSM
Beam 1
NLO laser
NLO laser long wavelength range
000000-2071-464
Beam 2
OPO
NLO laser short wavelength range
ZEISS
7

Hide quick links: