CHAPTER 1 - SYSTEM OPERATION
ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
The icons are not displayed, because the EFs are part of the filter cube together with the SBS filter, which
has to be replaced manually. The filter are indicated to SW within the Filters Tool window (Maintain tab).
The spectral detection band is displayed, and the type of filter is indicated in the tool tip for each
detector.
BiG.2 can be combined with up to 4 NDD PMT channels. The Light Path only shows the GaAsP channels
of BiG.2.
The Collimators... button opens the Adjust pinhole and collimator panel (Fig. 263). Use the slider,
the arrows or the input box to adjust the collimator position of the infrared light path. Optimize the
settings for maximum CPM (Counts per Molecule) rate.
Fig. 263
Adjust pinhole and collimator panel