Maintain; Set Spline - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

LSM 880
4.5.2

Maintain

The Maintain menu from the Menu bar contains
additional modules to check and guarantee the
interference-free operation of all the software and
hardware components of the LSM 710 and LSM
780 systems. In the Menu bar, click Maintain.
This opens the Maintain list (Fig. 22).
The maintain list is bipartite. On upper section lists
user accessible tools, the lower section lists tools
reserved for service personnel, they are password
protected.
See also section Maintain Tab for further
information.
4.5.2.1

Set Spline

This function permits calibration of the Scanner
position signals. This is required for the use of
spline curves in the Line scanning mode.
• Click on Set Spline in the Maintain list.
− The Spline window appears on the screen
(see Fig. 23).
• You can set the required Amplitude X,
Amplitude Y, Offset X and Offset Y values by
using the sliders.
• Click the Close button to leave the window.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Introduction to the Software Application Layout
000000-2071-464
Fig. 22
Maintain list
Fig. 23
Spline window