Zeiss LSM 880 Operating Manual page 521

Table of Contents

Advertisement

LSM 880
Center Screen Area / Image Containers - Display and Image Analysis
For the Diagram and Table, the X and Y pull down
menu and the Draw symmetric curve check box
will be additionally available. In the X and Y pull
down menu you can choose if you want to display
correlations in both (x and y) or only one (x or y)
direction (see Fig. 718 and Fig. 719).
If the Draw symmetric curve check box is
checked (Fig. 719), the correlation will be displayed
symmetrically around the zero spatial lag with
negative and positive lag values displayed. If the
box is unchecked, only the positive lag values will
be displayed.
For the Diagram only the Settings popup menu is
available (see (Fig. 720),. Here you can assign for
the X and Y Dimensions individual colors to the
Correlation, Fit results and Residuals diagrams
via the corresponding color palette drop down
menus. You can also define for each dimension the
line width by entering a number in the Line width
display boxes or using the arrows. By checking
the left and right Auto-Scale check boxes, the
diagrams for Correlation & Fit results as well as the
diagram for the Residuals, respectively, will be auto
scaled to the lowest and highest value. By de-
checking you can enter your own minimum and
maximum values into the corresponding Min. and
Max. display boxes by typing or by using the
arrows. All values between the minimum and
maximum will be scaled accordingly. Values lying
outside will not be displayed.
If the Center region box is checked the correlation will be centered on the pixel numbers that can be
put into the displayed Pixels input box. This allows you to display only the relevant part of the correlation
and avoid areas of no content.
If the Draw scaling bar box is checked, a Scaling will appear for any correlation diagram and deviated
diagram (Fit, Residuals, Map). For the intensity image, no scaling bar is available.
If the Use image palette box is checked, the look up table (LUT) selected for the intensity image will be
taken for the correlation diagram and Maps; the LUT will not apply for Fit and Residuals.
The Interpolation drop down menu (Fig. 721)
allows you to select, which data points will be
interpolated in the correlation function. If No
interpolation is selected, none of the data points
will be interpolated. If Interpolate G(0) is selected
the
zero
spatial
correlation
interpolated. If Interpolate G(0) and G(1) is
selected the zero spatial correlation and the 1
pixel shifted correlation value will be interpolated.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
value
will
be
st
000000-2071-464
Fig. 718
X and Y pull down menu
Fig. 719
Draw symmetric curve check box
Fig. 720
Settings popup menu
Fig. 721
Interpolation drop down menu
ZEISS
515