With a click on the Reuse button, the configuration used to record the selected measurement will be
loaded.
Fig. 363
Correlate window
Fig. 364
Correlate Error window
Automatic dust filter
If the Automatic cut box is checked, you can type in a threshold in % in the Automatic cut ratio
display box. This sets the threshold intensity for an electronic dust filter. The value (0 ... 100) will define
the threshold in percentage. A default value of 10 % can be selected by pressing the button on the right.
272
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
Reload
With Reload the current active measurement will
be reopened. If raw data have been stored and
attached to the measurement the Correlate
window will appear (Fig. 363).
You can redefine the following settings:
Automatic dust filter
Correlator settings
Count rate settings
Photon count histogram settings
Only
those
recalculated with the new settings which
have been selected in the original table. All
other
measurements
unchanged. To select all measurements use
the table context menu opened with a right
mouse click in the table and choose the line
Select all.
If no raw data have been stored or are
associated
with
measurement, an error message will be
displayed (Fig. 364). No manipulation of the
data is possible. Click Ok to continue
working with the unmodified data set.
LSM 880
measurements
will
will
be
opened
the
.fcs
file
or
10/2014 V_01