LSM 880
The Microscope Control tool contains the
following functions:
HAL On/Off button for transmitted light: If the
button is pressed, the control panel appears and
the shutter automatically opens (Fig. 353). If you
press the On button, the light intensity of the HAL
lamp can be varied via the input box or slider. The
light path will be highlighted by color in the
diagram. Press the On button again to shut down
the light. The color temperature for photo
documentation can be switched on via 3200 K
button.
HAL transmitted light Open/Close shutter
button: The shutter for transmitted light from the
HAL lamp is switched on and off by repeated
pressing.
Condenser button: The numerical aperture of the
condenser is set via input box or slider. The turret
position can be selected from graphical pop-up
menu (only for motorized condensers).
Stage button: If the stage button is pressed, the
Stage control panel will appear (Fig. 354). By
pressing the Stop button you can interrupt any
stage movement. The Move to Zero button will
move the stage to the current zero positions of the
x and y coordinates. The compass rose will move
the stage in the direction of the arrowheads. Fast
movement is obtained with the outer rose, slow
movement with the inner rose. The X-Position
(µm) and Y-Position (µm) display boxes show the
actual x and y coordinates. You can edit the
numbers by typing or using the arrowheads. If
numbers were altered, the first click on the
compass rose will bring the stage to the altered
position. The Step Size (µm) display box show the
currently use step size for stage movement. The
value can be edited by typing or using the
arrowheads. The Zero Position Set button set sets
the x and y coordinates to 0.
Focus button: If the focus button is pressed, the Focus control panel will appear (Fig. 355). The
Z-Position display box shows the current z-position. The number can be edited by typing or using the
arrowheads. The nosepiece or stage will automatically update the new position. By pressing the Stop
button, you can stop the current movement. The Step Size (µm) display box show the currently use step
size for stage movement. The value can be edited by typing or using the arrowheads. The Home button
sets the z-position to the predefined 0 value. Pressing Load will lower the nose piece (z-position) for
objective change. Work brings the z-position back to the actual used position if the focus was zeroed or
in Load position.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
Fig. 353
Locate - Microscope Control tool
Fig. 354
Locate - Microscope Control tool,
Stage
Fig. 355
Locate - Microscope Control tool,
Focus
ZEISS
265