LSM 880
6

Piezo Focusing Device

A Z-Piezo allows Z-stacks to be acquired considerably quicker compared to the focus drive of the
microscope stand.
The speed advantage depends on the total travel range in Z and the speed of the acquisition of the
individual image of a Z-stack.
The available travel range of a Z-Piezo device is much smaller compared to the Z drive of the microscope.
A Piezo device allows performing smaller steps in Z with a higher absolute precision for the individual
steps.
A Piezo device is recommended for 3D measurements of objects and for 3D superresolution imaging
(Airyscan).
Piezo Objective Focusing Device
For Axio Imager.Z2, Axio Imager.M2 and Axio Observer.Z1
Range:
Minimum step size:
Speed:
Slices
20
10/2014 V_01
CHAPTER 4 - ANNEX
Piezo Focusing Device
≥ 200 µm
15 nm
Step size [µm]
0.5
000000-2071-464
Piezo objective focusing device
xz-lines / s
460
ZEISS