Switching on and off

Before switching on the microscope system the movement range of the mechanical / measuring
stage must be cleared and any objects removed from the stage to prevent collisions and
damage during stage calibration.
The mechanical / measuring stage S mot. is switched on and off with the microscope and initializes
itself automatically when switched on.
After switching on the warning message Stage Calibration appears on the display of the
SYCOP 3 with the request to remove objects from the stage. Calibration can be started
immediately with OK or interrupted with Cancel and continued later as necessary. See
separate SYCOP 3 operating manual under:
Home\Microscope\XYZ Position\Function\auto) or
Home\Setup\Mode\Stage\Calibrate).
Calibration is necessary for use of the SYCOP 3 memory function. It is recommended that this
is carried out immediately upon prompting. The settings in the Memory menu are thus
available again after switching the system off/on. Calibration of the measuring stage is a
prerequisite for positional accuracy.
4.2
Control elements on the mobile
coaxial drive
XY coaxial drive for stage control
− Upper drive knob: Y adjustment (Fig. 11/5)
− Lower drive knob: X adjustment (Fig. 11/4)
− The drive knobs are height adjustable. Simply
push the drive knob up or down.
− If required, the friction of the drive knobs can
also be adjusted:
To do this, hold the black sleeve and slide the
silver sleeve in the corresponding direction to
increase or decrease friction.
− The horizontal position of the drive axis
(Fig. 11/6) can be individually selected:
this, loosen the two screws on the left
underside of the retaining bracket (Fig. 11/7)
and push the drive forwards or backwards.
Tighten the screws again.
08/2012
OPERATION
Switching on and off
To do
Fig. 11
435465-9000-701
Control elements on the mobile
coaxial drive
Carl Zeiss
21