The global link status of a parameter is displayed in the Link column. By clicking with the left mouse
button in the respective field, a button tab becomes available. Use the arrow head to open the drop
down menu. Five options for the linking a parameter will become available (Fig. 384):
Fig. 384
− Link measurement: links a parameter from different measurements (abbreviated M)
− Link kinetic indexes: links a parameter from the same time points (abbreviated K)
− Link positions: links a parameter obtained from measurements at the same site (abbreviated K)
− Link repetitions: links a parameter for all repetitions of one measurement (abbreviated R)
− Link channels: link a parameter for the same channels (abbreviated C)
You can select the type of linkage by checking the corresponding boxes. More than one link can be
activated. After closing the button tab, the linkage type is displayed as a one letter abbreviation. If a
linkage is activated that does not apply for a measurement, it will be just disregarded. Global fitting will
fit all linked parameters of different measurements to the same value. All repetitions regardless of being
selected or not in the result table will be taken for global fitting. If you want to exclude measurements
from the global fit, they have to be deleted from the result table.
• Select one or more rows in the Fit Table, for which you want to perform the fit. All curves of the
selected row will be displayed. The last selected row will determine the channel that will be displayed
in the Channel display field.
PCH can be fitted only one-dimensional. If a cross-correlation measurement is activated, the PCH
model will be automatically replaced by a correlation model.
• Set the range of the curves to be fitted by setting the red and blue bars in the correlation diagram
(PCH diagrams do not have this option and the whole data set is always fitted). Press the Fit button to
fit the loaded curves to the model. If you press Fit all all measurements with the same channel as the
one displayed in the Channel field will be fitted as well.
All other channels will not be fitted and the model has to be selected anew, if such a row with a
different channel is selected.
By pressing the Define... button (or selecting the Define model from the context menu of the Fit
table) the Define model window will open with the same functionalities as described in the Set up Fit
toolbar.
288
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
Link drop down menu