Center Screen Area / Image Containers - Display and Image Analysis
6.5.3
Airyscan – View Option Tab Airyscan
2D Processing Image preview
There are several options available for the display of the Airyscan data.
Source image
Source image (Fig. 619) shows the raw sum data
useful for focusing, panning, and setting of laser
power and detection gain.
SR (VP) image
SR (VP) image gives a real time impression of the
2D processing result (Fig. 620). This can be both
the SR mode and the VP mode. You can see the
improved resolution or the modified sectioning this
way.
Deselecting the checkboxes will hide the respective
image.
Fig. 620
2D processing result
Fig. 621
Intensity of the Airy orders
In addition, there are some intensity displays available. The table shows the intensity of the Airy orders
(Fig. 621).
Note that in case of inadequate sample quality or acquisition settings (e.g. overexposure) the
typical drop of intensities from low to high ring numbers is not available or reversed.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Fig. 619
000000-2071-464
Airyscan View – Airyscan tab