Courses On "How To Operate The System In An Optimized Way; Beam Path Configuration Guide For Fluorescence Microscopy; Optical Elements In The Configuration Control Window; Main Dichroic Beam Splitter (Mbs) - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

LSM 880
2

Courses on "How to Operate the System in an Optimized Way"

ZEISS is offering training courses on how to operate the system in an optimized way.
− Courses are held in our application center in Jena, Germany.
− Courses are held in English or German language, respectively.
Check out:
www.zeiss.de/lsm
for the latest dates and ask your Zeiss representative for a quotation on courses.
3

Beam Path Configuration Guide for Fluorescence Microscopy

3.1

Optical Elements in the Configuration Control Window

All wavelength values are given in Nanometer [nm].
3.1.1

Main Dichroic Beam Splitter (MBS)

• A MBS XXX/YYY deflects the indicated laser lines onto the specimen and allows the emitted
fluorescent light to pass through.
Example: MBS 458/514, MBS 488/543/633
• A MBS XXX+ is a special type of a main dichroic used for IR multiphoton excitation. The MBS 690+
for example, deflects laser light with a wavelength longer as 690 nm onto the specimen and allows
fluorescent emission light in the visible range below 660 nm to pass through.
Example: MBS 690+
• A MBS XXX/YYY+ is a combination of a MBS XXX and MBS YYY+ used for simultaneous IR
multiphoton and single photon excitation.
Example: MBS 405/710+.
10/2014 V_01
CHAPTER 4 - ANNEX
Courses on "How to Operate the System ...
000000-2071-464
ZEISS
5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lsm 880 nlo

Table of Contents