Maintain Menu; Scanner; Electrical Calibration - Zeiss LSM 510 Manual

Hide thumbs Also See for LSM 510:
Table of Contents

Advertisement

Carl Zeiss
4.10

Maintain Menu

The Maintain menu contains additional modules to check and guarantee the interference-free operation
of all the software and hardware components of the LSM 510. The LSM 510 and LSM 510 META can be
equipped with an additional scanner, the LSM DuoScan. For optimal results the scan fields of the two
scanners must be aligned. This procedure is described in chapter 4.8.4.1
• In the Main menu toolbar, click on Maintain. This opens another subordinate toolbar in the Main
menu.
Fig. 4-223
Maintain menu
Fig. 4-224
Scanner Calibration window
4.10.1.1

Electrical Calibration

Electrical Calibration with Speed 1-10 can be performed unidirectional / bidirectional, but with
Speed 11-13 only unidirectional
(1)
Preliminary notes
The electrical calibration has to be performed every 2-3 months. For electrical calibration no laser
scanning is performed and for that reason no calibration sample is needed.
4-226
OPERATION IN EXPERT MODE
Maintain Menu
4.10.1
The Scanner function is used for scanner
calibration.
The following type of calibrations are available:
− Electrical
− Electrical calibration with Speed 11-13
− Optical
B 45-0021 e
LSM 510 META DuoScan

Scanner

calibration
(unidirectional / bidirectional)
(unidirectional)
calibration
(bidirectional)
LSM 510 DuoScan
with
Speed
1-10
with
Speed
11-13
03/06