ALGORITHM CALIBRATION
PM_yld
D
Ill. 6
A
B
LNRT PARAM display x=(3 values)
CHANGE
LNRT PARAM change x=(value)
NEXT
VALID
C
TIME RECIPROCITY FAILURE THICK.=x cm
CALIB
TIME RECIPROCITY FAILURE
D
CALIB REF_ENERGY OF LNRT
S2100D How To.fm
E
LNRT
Energ
Corr
D
Ill. 6
MAGNIFICATION CORRECTION
CALIB
MAGNIFICATION CALIB. Config:contact/magnify
CONFIG
CONTACT CONFIGURATION
YES
NO
OD_mea
CALC
RECIPROCITY FAIL.:calc
VALID
MEASURED OPTICAL DENSITY=(value)
NEXT
VALID
CALIB REF_ENERGY OF LNRT mAs=(value)
CALIB
Service Information and Procedures Class A 2385072-16-8EN
LNRT
(TIME RECIP. FAIL.)
PARAM
CALIB
CALC
MAGNIFICATION CALIB calc
VALID
Select plexi thickness
2 CM
OD_mea
CALC
MEASURED DENSITY OD=(value)
NEXT
VALID
Page no. 113
Senographe Operations - How to...
(GENE)
ALGO
MAG.
CORR.
MAGNIF. PARAM display x=(2 values)
CHANGE
MAGNIF. PARAM change x=(value)
NEXT
VALID
MAGNIFICATION CALIB. x cm
THICK
CALIB
MAGNIFICATION CALIB.
4 CM
6 CM
CALIB REF. ENERGY calc
VALID
Senographe DS
Chapter 3