GE Senographe DS Manual page 108

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Senographe Operations - How to...
ILLUSTRATION 5 -
FREQUENCY/PM_HV GAIN
CALIB
FREQUENCY/PM_HV GAIN ref=(value)
REF
Send PM/HV ref
START
FREQ/PM_HV REF modif ref=(value)
NEXT
VALID
default=+7.000E+2
Chapter 3
PARAM
FREQ/PM_HV GAIN display F/HV=(value)
CHANGE
FREQ/PM_HV GAIN change F/HV=(value)
NEXT
VALID
CALIB
HV_M
CALC
FREQ/PM_HV calcul
STOP
FREQ/PM_HV Measure meas=(value)
NEXT
Service Information and Procedures Class A 2385072-16-8EN
PM_HV/DAC CALIBRATION
FREQ/HV
PM_HV/DAC Calibration
START
PM_HV/DAC PARAM display alpha=(2 values)
CHANGE
PM_HV/DAC PARAM change alpha=(value)
NEXT
VALID
VALID
Page no. 108
HV/DAC
PM_HV/DAC Calibration
CALIB
PARAM
VALID
Senographe DS
A
S2100D How To.fm