GENERATOR CALIBRATION
TUBE
HTR_SCL
A
Ill. 3
A
kV CALIBRATION
CALIB
kV CALIBRATION
1st pt
2nd pt
Same menu
structure as for
1st pt
kV REFERENCE 1st point ref=(value) kV
NEXT
VALID
S2100D How To.fm
BIAS
KV
mA_meas
PARAM
kV PARAMETERS x=(2 values)
CHANGE
PARAM kV MODIF x=(value)
NEXT
VALID
calcul
THIS MENU IS ONLY ACCESSIBLE
WHEN BOTH HV VALUES HAVE
BEEN MEASURED AND ENTERED
VIA THE "1ST PT" AND "2ND PT"
MENUS.
VALID
kV CALIB 1st point ref=(value) kV
REF
CALIB
kV CALIB 1st point
Service Information and Procedures Class A 2385072-16-8EN
BIAS
mA MEASUREMENT
CALIB
PARAM
mA MEASURE CALIBRATION
CALIB
I_MEAS
CALCUL
mA MEASUREMENT CALC
MEASURED CURRENT meas=(value)
NEXT
VALID
mA MEASURE CALIBRATION meas=x mA
MEASUR
PRESSING THE "MEASUR" KEY UP-
DATES THE DISPLAYED VALUE OF THE
THEORETICAL MEASURED X-RAY
TUBE CURRENT VALUE
kV_M
THIS MENU IS ONLY
ACCESSIBLE WHEN 1ST PT
KV VALUE HAS BEEN
MEASURED VIA THE "CALIB"
MENU.
MEASURED kV 1st meas=(value)
NEXT
VALID
Page no. 107
Senographe Operations - How to...
GENE
kV
mA_meas
THIS MENU IS ONLY
ACCESSIBLE WHEN X-RAY
TUBE CURRENT HAS BEEN
MEASURED AND ENTERED
VIA THE "CALIB" AND
"I_MEAS" MENUS.
VALID
mA MEASUREMENT PARAM f/mA=(value)
CHANGE
mA MEAS PARAM MODIF f/mA=(value)
NEXT
VALID