GE Healthcare
Revision 1
1-9. The DISP Parameter
The DISP (display) parameter is provided to aid service procedures; it controls the display of certain
parameters after an exposure. When it is set to YES, the two groups of exposure parameters shown
in the table below are displayed alternately
Group 1:
Group 2:
Composition
Composition
Thickness:
Prep dose:
Prep kV:
Prep track:
Prep filter:
Delta grid:
To change the DISP parameter, access SETUP/GENE/INSTAL/GENE/CONFIG/DISP, and select YES
or NO. A CKSUM is NOT necessary because this parameter is not a saved parameter. When power is
turned off, the value of DISP is lost and reverts to NO at the next power-up.
S2100D How To.fm
kV
mAs
Thickness
Breast composition
Breast thickness in equivalent cm of acrylic.
Dose for pre-shot (returned by IDC)
kV used for pre-shot
Track used for pre-shot
Filter used for pre-shot
Delta time (ms) from end of X-ray emission to grid U-turn
Service Information and Procedures Class A 2385072-16-8EN
:
mA
Time
Prep dose
Prep kV
Page no. 97
Senographe Operations - How to...
Ih
Prep track
Prep filter
Senographe DS
Delta grid
Chapter 3