Revision 1
Product history and compatibility
3-1-4
Nephtys M410 (IS4-3 FMI 12075)
The AWS Application Software for the V3 Control Stations (U20 Workstation) was updated from
V53.0 to V53.10.1.
The AWS Application Software for the V1/V2 Control Stations (SB150 Workstation) was updated
from V32.10 to V43.10.1.
The Gantry CPU board was updated from NSC.1.0.10 to NSC.1.1.7.
The Interface board was updated from FPGA 2E to FPGA 4A.
FMI 12075 (DS) installed on the install base.
3-1-5
Nephtys M411 (IS4-3 Fwd Prod)
Introduction of Sisfle Rotative Arm (replaces Ergotron B Rotative Arm) on the Control Station.
The Gantry CPU board for manufacturing was updated from NSC.1.1.7 (5233827-2) to NSC.1.1.8
(5233827-3).
Introduction of V2 Clutch, which is only compatible with the Rotation Board (2375488-12). However,
the Rotation Board (2375488-12) is compatible with both V1 and V2 Clutches.
Note:
The Rotation Board (2375488-11) is not a FRU. If you need to replace the Rotation Board
(2375488-11), order Rotation Board (2375488-12).
A new Gantry Update CD-ROM (5174994-7) is delivered with the system.
Note:
The Gantry Update CD-ROM (5174994-7) is not a FRU. If you need to replace the Gantry CD-
ROM, order Gantry Update CD-ROM (5174994-6).
3-1-6
Nephtys M412 (Orion 1)
The support of 5 new languages and the replacement of Brazilian Portugese by continental Portugese,
which results in the introduction of:
a new Generator CPU Board (2385602-9), which supports additional languages.
a new Gantry CPU Board (5233827-4), which supports additional languages.
ADS 53.10.10 on V3 Control Stations, which supports 11 languages in the user interface, and 19 lan-
guages in the embedded user documentation.
a new Lift Board (2373567-9), which supports output of additional characters to the Gantry LCD
Readout.
Other product enhancements, including the introduction of:
a new Paddle Holder Board (2369391-2), which has DIP switches and better reliability. The Paddle
Holder Board (2369391-2) is fully backward comptible.
new Detector FRUs (2383172-4 and 2383172-5).
Type 2 Lift Screw Assembly. The Type 2 Lift Screw Assembly differs from the Type 1 Lift Screw
Assembly, such that it uses a different Ball Bearing Ring, and its associated Lift Motor has an earth
cable.
3-1-7
Nephtys M412 Step1
Introduction of X-ray Console (2282675-7) and new Arm Keypads (5268425).
3-1-8
Nephtys M413
Introduction of one Gantry board FRU strategy.
Safety compliance : New tools and filmless procedure available for Collimator checks.
Introduction of a stronger 3000 N Rotation Gas Spring (5154719).
Introduction of new Arm Keypads (5268425-2).
Chapter 3
Service Information and Procedures Class A 2385072-16-8EN
Page no. 60