• Background: Update On, Display Off (this allows a trace to be blanked and continue to update "in the
background")
Key Path
Notes
Couplings
Readback Text
Initial S/W Revision

Trace Update State (Remote Command Only)

Toggles the trace update state.
Remote Command
Example
Couplings
Preset
State Saved
Initial S/W Revision
Trace Display State (Remote Command Only)
Toggle the trace display state.
Remote Command
EMI Receiver Mode Reference
Trace/Detector
The four states of this 1-of-N actually set two variables, Update and Display, to their four possible
combinations:
Trace On: Update and Display both On
View: Update Off and Display On
Blank: Update Off and Display Off
Background: Update On, Display Off
See the following tables,
"Trace Update State (Remote Command Only)" on page 581
Display State (Remote Command Only)" on page
variables.
Selecting a trace type (Clear Write, Trace Average, Max Hold, Min Hold) for a trace (pressing the key
or sending the equivalent remote command) puts the trace in Trace On (Update On and Display On),
even if that trace type was already selected.
Selecting a detector for a trace (pressing the key or sending a [:SENS]:DET:TRAC command) puts the
trace in Trace On (Update On and Display On), even if that detector was already selected.
Loading a trace from a file puts that trace in View regardless of the state it was in when it was saved;
as does being the target of a Copy or a participant in an Exchange.
When a View/ Blank selection is made, the menu returns to the previous menu.
Trace On|View|Blank|Background
A.07.00
:TRACe[1]|2|3:FSCan:UPDate ON|OFF|1|0
:TRACe[1]|2|3:FSCan:UPDate?
FSC:TRAC2:UPD 0  Makes trace 2 inactive (stops updating).
FSC:TRAC2:UPD  Query update state of trace 2.
Whenever you set Update to On for any trace, the Display is set to On for that trace.
1|0|0 (On for Trace 1; Off for 2–3)
Saved in instrument state
A.07.00
:TRACe[1]|2|3:FSCan:DISPlay ON|OFF|1|0
8  Frequency Scan Measurement
581, for details on the SCPI to control these two
Trace/ Detector
and
"Trace
581