Trace Fail? (Remote Command Only) - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

SCPI Notes
Initial S/W Revision

Trace Fail? (Remote Command Only)

Tests a trace against all associated limit lines. Returns a 0 if the trace is within all limits and margins, a 1 if
the trace exceed either the limit or the margin. If no limits apply to the selected trace, this will
automatically return a 0.
Only applies to limits that are turned on, if a Limit is off it will not be tested.  If a Trace is not displaying it will
still be tested, and if Test Limits (All Limits) is off the Trace will still be tested.
This command ignores limit lines that are assigned to other traces.
Remote Command
Example
Initial S/W Revision
Limit Line Control (Remote Command Only)
Defines a list of limit line control (frequency or time) values for a given limit line. Up to 2000 points may be
defined for each limit using the following parameters.
<x> Frequency or time values as specified by :Calculate:LLINe:CONTrol:DOMain. Units default to Hz (for
frequency) and seconds (for time).
Range: –30 Gs to +30 Gs for time limits, –3 kHz to +1200 GHz for frequency limits.
Remote Command
Example
Preset
State Saved
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision
Remote Command
EMI Receiver Mode Reference
measurement, they are added solely for remote language compatibility.
A.13.00
:CALCulate:MONitor:TRACe[1]|2|3:FAIL?
CALC:MON:TRAC3:FAIL? returns a zero if there is no failure, 1 if the trace exceeds either the limit or
the margin.
A.13.00
:CALCulate:LLINe[1]|2|...|6:CONTrol[:DATA] <x>, <x>, ...
:CALCulate:LLINe[1]|2|...|6:CONTrol[:DATA]?
CALC:LLIN:CONT 1GHz, 2GHz, 2GHz, 3GHz  Describes the X values of a stair-stepped limit line.
CALC:LLIN:CONT?
Unaffected by Preset
Saved in instrument state
:CALCulate[1]|2:LIMit[1]|2|...|6:CONTrol[:DATA]
This command above is included for ESU compatibility. The suffix 1|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
A.13.00
:CALCulate:LLINe[1]|2|...|6:CONTrol:POINts?
10  Monitor Spectrum Measurement
Meas Setup
803

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents