Internet Explorer - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

SYSTem:COMMunicate:LAN:DGATeway?
Subnet
SYSTem:COMMunicate:LAN:SMASk <string>
Mask
SYSTem:COMMunicate:LAN:SMASk?

Internet Explorer...

This key launches Microsoft Internet Explorer.  A mouse and external keyboard are highly desired for using
Internet Explorer.  When Internet Explorer is running, close Internet Explorer to return focus to the
Instrument Application (or use Alt-Tab).
This feature is not available if option SF1 is installed.
Key Path
Mode
Notes
Initial S/W Revision
System Remote Commands (Remote Commands Only)
The commands in this section have no front-panel key equivalent.
"System Powerdown (Remote Command Only)" on page 353
"List installed Options (Remote Command Only)" on page 354
"Lock the Front-panel keys (Remote Command Only)" on page 354
"List SCPI Commands (Remote Command Only)" on page 354
"SCPI Version Query (Remote Command Only)" on page 355
"Date (Remote Command Only)" on page 355
"Time (Remote Command Only)" on page 355
Initial S/W Revision
System Powerdown (Remote Command Only)
Remote Command
Notes
EMI Receiver Mode Reference
System
All
No equivalent remote command for this key.
A.05.01
Prior to A.02.00
SYSTem:PDOWn [NORMal|FORCe]
Shuts down the instrument in the normal way (NORMal) or forced way (FORCe). In case there is
another application with modified data pending for saving, the application prompt the user. The
system waits until the user responds in the normal mode. It will go off after 20 seconds of wait in the
force mode and all data will be lost.
:SYSTem:COMMunicate:LAN[:SELF]:GATEway?
:SYSTem:COMMunicate:LAN[:SELF]:SUBNetmask <string>
:SYSTem:COMMunicate:LAN[:SELF]:SUBNetmask?
6  System Functions
System
353

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents