Trace/Detector

Displays a menu of keys that enable you to control the trace setting for the APD trace window.
Key Path
Notes
Initial S/W Revision
Select Trace
Determines which trace the type control keys will affect. When you select a trace, it makes that trace the
current trace, so it displays on top of all of the other traces.
Key Path
Notes
Preset
State Saved
Initial S/W Revision
View/Blank
This key lets you set the state of the two trace variables, Update and Display. The four choices available in
this 1-of-N menu are:
• Trace On: Update and Display both On
• View: Update Off and Display On
• Blank: Update Off and Display Off
• Background: Update On, Display Off (this allows a trace to be blanked and continue to update "in the
background")
Key Path
Notes
Couplings
EMI Receiver Mode Reference
Front-panel key
Front panel only.
A.13.00
Trace/ Detector
Front panel only.
Measured
Saved in instrument state.
A.13.00
Trace/Detector
The four states of this 1-of-N actually set two variables, Update and Display, to their four possible
combinations:
Trace On: Update and Display both On
View: Update Off and Display On
Blank: Update Off and Display Off
Background: Update On, Display Off
See the following tables,
"Trace Update State (Remote Command Only)" on page 1010
Display State (Remote Command Only)" on page
variables.
Loading a trace from a file puts that trace in View regardless of the state it was in when it was saved.
1010, for details on the SCPI to control these two