Clear List And Start - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Couplings
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision

Clear List and Start

This key will perform the following when pressed:
1. Prompt you for confirmation to delete all signals in signal list with the following message:
2. After confirmation, performs the selected scan sequence. This will turn the label on the Start key to
Stop. If the selected Scan Sequence is Scan, and the scan is set to continuous, you need to stop the
measurement with the Stop key. For all other cases, the scan sequence will stop when the measurement
has completed.
Key Path
Remote Command
Example
Dependencies
Initial S/W Revision
Single (Meters)
Performs a single measurement at meters.
Key Path
EMI Receiver Mode Reference
When stopped, the label on the key changes to Start.
:ABORt
This SCPI is remote language compatible with ESU.
A.07.00
Sweep/Control
INITiate:FSCan:CLEar:IMMediate
INIT:FSC:CLE:IMM
This key is disabled in two conditions:
1.A Scan sequence is running. If the grayed out key is pressed, an advisory message is generated. If
the equivalent SCPI command is sent, this same message is generated as part of a "–221, Settings
conflict" warning.
2.A Scan Sequence is set to Remeasure. If the grayed out key is pressed, an advisory message is
generated. If the equivalent SCPI command is sent, this same message is generated as part of a "–
221, Settings conflict" warning.
A.07.00
Sweep/Control
8  Frequency Scan Measurement
Sweep/ Control
575

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents