4  Input/Output Functions
Input/Output
Mode
Dependencies
Preset
Initial S/W Revision
Modified at S/W Revision

Select Correction

Specifies the selected correction. The term "selected correction" is used throughout this document to
specify which correction will be affected by the functions.
Key Path
Mode
Notes
Preset
Readback
Initial S/W Revision
Modified at S/W Revision
Correction On/Off
Turning the Selected Correction from the OFF state to the ON state allows the values in it to be applied to
the data. This state transition also automatically turns on "Apply Corrections" (sets it to ON), otherwise the
correction would not take effect.
A new sweep is initiated if an amplitude correction is switched on or off.  Note that changing, sending or
loading corrections data does NOT directly initiate a sweep, however in general these operations will turn
corrections on, which DOES initiate a sweep.
Key Path
Remote Command
184
SA, I/Q Analyzer, Phase Noise, VXA, RTSA, EMI Receiver, DVB-T/H, DTMB, DVB-T/H, DTMB, W-
CDMA, LTE & LTE-Adv FDD, LTE & LTE-Adv TDD, Sequence Analyzer, BTooth
This key will only appear if you have the proper option installed in your instrument.
Amplitude correction may not be available in all modes; if a mode does not support amplitude
correction, the Corrections key should be blanked while in that mode.  If an application supports
corrections but the current measurement does not, then the key should be grayed out in that
measurement
Corrections arrays are reset (deleted) by Restore Input/Output Defaults.  They survive shutdown and
restarting of the analyzer application, which means they will survive a power cycle.
A.02.00
x.14.50
Input/Output, Corrections
SA, I/Q Analyzer, Phase Noise, VXA, RTSA, EMI Receiver, DVB-T/H, DTMB, DVB-T/H, DTMB, W-
CDMA, LTE & LTE-Adv FDD, LTE & LTE-Adv TDD, Sequence Analyzer, BTooth
The selected correction is remembered even when not in the correction menu.
Set to Correction 1 by Restore Input/Output Defaults
Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction
6|Correction7|Correction8
A.02.00
x.14.50
Input/Output, Corrections
[:SENSe]:CORRection:CSET[1]|2|...|8[:STATe] ON|OFF|1|0