Delete All

Marks and deletes all signals in the signal list. If you accessed  this function from the front panel, the
following message will prompt you  for confirmation to delete the selected signals from the signal list..
This will delete all signals.  Are you sure you want to do this?
Please press Enter or OK to proceed. Press ESC or Cancel to close this dialog.
Key Path
Remote Command
Example
Notes
Initial S/W Revision
Delete Marked
Deletes all marked signals in the signal list. If you accessed  this function from the front panel, the
following message will prompt you  for confirmation to delete the selected signals from the signal list.
This will delete the marked signals. Are you sure you want to do this?
Please press Enter or OK to proceed. Press ESC or Cancel to close this dialog.
Key Path
Remote Command
Example
Notes
Initial S/W Revision
EMI Receiver Mode Reference
Meas Setup, Signal List, Delete Signals
:CALCulate:SLISt:DELete:SIGNal <integer>
CALC:SLIS:DEL:SIGN 4  Delete signal #4.
If the signal list contains no data, an error is generated and this function cannot be performed.
When signals in the signal list were successfully deleted, an advisory is generated.
If the selected signal is not in the list, an error is generated and this function cannot be performed.
A.13.00
Meas Setup, Signal List, Delete Signals
:CALCulate:SLISt:DELete:ALL
CALC:SLIS:DEL:ALL
If the signal list contains no data, an error is generated and this function cannot be performed.
When signals in the signal list were successfully deleted, an advisory is generated.
A.13.00
Meas Setup, Signal List, Delete Signals
:CALCulate:SLISt:DELete:MARKed
CALC:SLIS:DEL:MARK
If the signal list contains no data, an error is generated and this function cannot be performed.
If no marked signal is found in the signal list, an error is generated and Delete Marked cannot be
performed.
When signals in the signal list were successfully deleted, an advisory is generated.
A.13.00