Autorange - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Notes
Dependencies
Preset
State Saved
Backwards
Compatibility SCPI
Initial S/W Revision

Autorange

When autorange is set to on, the attenuation will be increased automatically when an overload is detected
during a scan.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision
Auto Preamp
When on, the Int Premp will be considered in the autorange procedure when performing a scan.
Key Path
EMI Receiver Mode Reference
SCAN5:INP:GAIN?
This command is included for ESU compatibility. The suffix [1]|2 of SENSe node is irrelevant to the
measurement, they are added solely for remote language compatibility.
ON is aliased to :SCAN:POW:GAIN ON; :SCAN:POW:GAIN:BAND LOW for the selected range.
OFF is aliased to :SCAN:POW:GAIN OFF for the selected range.
This key is grayed out when a scan sequence is running except during single range scanning. If the
grayed out key is pressed, an advisory message is generated. If the equivalent SCPI command is sent,
this same message is generated as part of a "–221, Settings conflict" warning.
OFF
Saved in instrument state.
:SENSe1|SENSe2:SCAN[1]|2|...|10:INPut:GAIN[:STATe]
A.07.00
Meas Setup, Scan Table
[:SENSe]:FSCan:SCAN[1]|2|...|10:INPut:ATTenuation:AUTO ON | OFF | 1 | 0
[:SENSe]:FSCan:SCAN[1]|2|...|10:INPut:ATTenuation:AUTO
:FSC:SCAN8:INP:ATT:AUTO 1  Turn on Auto range for Range 8
:FSC:SCAN8:INP:ATT:AUTO?
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
OFF
Saved in instrument state.
:SENSe1|SENSe2:SCAN[1]|2|...|10:INPut:ATTenuation:AUTO
[:SENSe]:SCAN[1]|2|...|10:INPut:ATTenuation:AUTO
This command above is included for ESU compatibility. The suffix [1]|2 and [1]|2|...|10 are irrelevant to
the measurement, they are added solely for remote language compatibility.
A.07.00
Meas Setup, Scan Table
8  Frequency Scan Measurement
Meas Setup
467