sweep point. The Step Size is calculated by this formula: Step Size = (Stop Frequency—Start Frequency)/
(Sweep Point-–1).
Selecting a MIL Preset will set the start/ stop frequencies, step size (in according to sweep point), and
RBW of the selected range according to the table values shown below. Point/RBW will be set to 2 for
Discrete and Smooth scan type.
30 Hz – 1
MIL Band
kHz
30 Hz
Start
Frequency
1 kHz
Stop
Frequency
10 Hz
RBW
195
Sweep Point
Dwell Time
150 ms
5 Hz
Step Size
The stop frequency of Band E  depends on the maximum frequency, with a maximum of 18 GHz. The
number of points in Band E is as below:
Option
503 (3.0 GHz Models)
503 (3.6 GHz Models)
507 (7 GHz Models)
507 (7.5 GHz Models)
508
513
526

Scan Sequence

Displays the menu keys that enable you to control the measurement sequence. You need to select the
Start key to start the selected measurement sequence. When Frequency Scan runs, some or all of the
following steps are included:
Scan:
If Scan is part of the current Scan Sequence, any hold or average trace will be reset. The Scan goes
through all of the range tables that are currently on, and scans the specified range using the range table
parameters. The detector(s) used for the scan are those specified in the Detectors menu under Trace/Det.
Each trace whose display and update state are not Off is displayed as it is being built. A maximum of 3
traces can be running simultaneously. All setting changes are locked during the scan, but you are still able
EMI Receiver Mode Reference
1 kHz – 10
10 kHz –
kHz
150 kHz
1 kHz
10 kHz
10 kHz
150 kHz
100 Hz
1 kHz
181
281
15 ms
15 ms
50 Hz
500 Hz
Max Freq
Sweep Points
3.0 GHz
4001
3.6 GHz
5201
7.0 GHz
12001
7.5 GHz
13001
8.4 GHz
14801
13.2 GHz
24401
18 GHz
34001
150 kHz –
30 MHz – 1
30 MHz
GHz
150 kHz
30 MHz
30 MHz
1 GHz
10 kHz
100 kHz
5971
19401
15 ms
15 ms
5 kHz
50 kHz
8  Frequency Scan Measurement
Meas Setup
Above 1 GHz
1GHz
Max freq of analyzer or 18
GHz, whichever is lower
1 MHz
Refer below.
15 ms
500 kHz
471