Application Identification Catalog (Remote Commands Only) - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Preset
State Saved
Initial S/W Revision

Application Identification Catalog (Remote Commands Only)

A catalog of the installed and licensed applications (Modes) can be queried for their identification.
"Application Catalog Number of Entries" on page 231
"Application Catalog Model Numbers" on page 231
"Application Catalog Revision" on page 231
"Application Catalog Options" on page 232
Application Catalog Number of Entries
Returns the number of installed and licensed applications (Modes).
Remote Command
Example
Preset
State Saved
Initial S/W Revision
Application Catalog Model Numbers
Returns a list of Model Numbers for the installed and licensed applications (Modes).
Remote Command
Example
Notes
Preset
State Saved
Initial S/W Revision
Application Catalog Revision
Returns the Revision of the provided Model Number.
EMI Receiver Mode Reference
Not affected by a Preset
Not saved in state per se, the value will be the selected application when a Save is invoked.
Prior to A.02.00
:SYSTem:APPLication:CATalog[:NAME]:COUNt?
:SYST:APPL:CAT:COUN?
Not affected by Preset
Not saved in instrument state.
Prior to A.02.00
:SYSTem:APPLication:CATalog[:NAME]?
:SYST:APPL:CAT?
Returned value is a quoted string of a comma separated list of Model Numbers. Example, if SAMS
and Phase Noise are installed and licensed:
"N9060A,N9068A"
String length is COUNt * 7 – 1. (7 = Model Number length + 1 for comma. –1 = no comma for the
1st entry.)
Not affected by a Preset
Not saved in instrument state.
Prior to A.02.00
5  Mode Functions
Mode
231

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents