Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Default Unit
Status Bits/OPC
dependencies
Initial S/W Revision
CF à Signal (Replace)
Replaces the frequency of the selected signal with the Center Frequency (and Meters Freq since they are
the same). Amplitude values and Delta To Limit values will be shown as "---" to indicate that those values
are undefined until the next scan/final measurement value.
Key Path
Remote Command
Example
Notes
Initial S/W Revision
CF à List (Append)
Append the signal at the Center Frequency (and Meters Freq since they are the same) into the Frequency
Scan signal list.  Amplitude values and Delta To Limit values will be shown as "---" to indicate that those
EMI Receiver Mode Reference
FREQ Channel
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
FREQ:CENT:STEP:AUTO ON
FREQ:CENT:STEP 500 MHz
FREQ:CENT UP increases the current center frequency value by 500 MHz
FREQ:CENT:STEP?
FREQ:CENT:STEP:AUTO?
Preset and Max values are dependent on Hardware Options (503, 508, 513, 526)
Auto
ON
Saved in instrument state
– ( the maximum frequency of the instrument). (that is, a 27 GHz max freq instrument has a CF step
range of +/– 27 GHz)
The maximum frequency of the instrument. (that is, a 27 GHz max freq instrument has a CF step
range of +/– 27 GHz)
Hz
non-overlapped
A.07.00
Freq
:CALCulate:MONitor:SLISt:REPLace:METer <integer>
CALC:MON:SLIS:REPL:MET 20 Replace Signal #20 with Center Freq
If the signal list contains no data, an error is generated and this function cannot be performed.
If the selected signal is not in the list, an error is generated and this function cannot be performed.
A.13.00
10  Monitor Spectrum Measurement