4 Input/Output Functions
Input/Output
Notes
Preset
State Saved
Range
Min
Max
Initial S/W Revision
Coupling
Some probe types allow coupling to reject low frequencies. This will filter out the DC component of a signal
that is composed of a DC bias plus some AC signal. This control is visible only for probe types that have this
capability.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Range
Readback Text
Initial S/W Revision
DC
Turns off low frequency rejection, allowing signals down to DC.
Key Path
Example
Initial S/W Revision
174
Only some probe types support Offset. For those that do, each probe type has its own Offset setting.
As probes are changed the Offset value will reflect the new probe's setting. Changing the Offset
affects only the current probe type's setting and leaves all others unchanged.
0 V
Saved with probe calibration data. It survives power cycle and is not affected by Preset or Restore.
–18 V to +18 V
–18 V
+18 V
Prior to A.02.00
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe
:INPut:COUPling:I|Q DC|LFR1|LFR2
:INPut:COUPling:I|Q?
Set the probe to low frequency rejection below 1.7 Hz.
INP:COUP:I LFR1
Only some probe types support Coupling. For those that do, each probe type has its own Coupling
setting. As probes are changed the Coupling value will reflect the new probe's setting. Changing the
Coupling affects only the current probe type's setting and leaves all others unchanged.
DC
Saved with probe calibration data. It survives a power cycle and is not affected by a Preset or
Restore.
DC | AC 1.7 Hz LFR1 | AC 0.14 Hz LFR2
DC | LFR1 | LFR2
Prior to A.02.00
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Turn off low frequency rejection on the I channel
INP:COUP:I DC
Prior to A.02.00
EMI Receiver Mode Reference