A trace can have multiple limit lines simultaneously; in that case, only one upper limit line will affect the
color of the trace. Other limit lines will be displayed, and will affect the pass/fail status, but the trace will
not turn red if it crosses a secondary limit line.
Key Path
Remote Command
Example
Notes
Couplings
Preset
State Saved
Range
Initial S/W Revision

Interpolation

Accesses a menu that lets you set the frequency and amplitude interpolation of the selected limit.
Key Path
Readback
Initial S/W Revision
Frequency Interpolation
Sets the interpolation between frequency points, allowing you to determine how limit trace values are
computed between points in a limit table. The available interpolation modes are linear and logarithmic. If
frequency interpolation is logarithmic (Log), frequency values between limit points are computed by first
taking the logarithm of both the table values and the intermediate value. A linear interpolation is then
performed in this logarithmic frequency space. An exactly analogous manipulation is done for logarithmic
amplitude interpolation.
Note that the native representation of amplitude is in dB.
For linear amplitude interpolation and linear frequency interpolation, the interpolation is computed as:
For linear amplitude interpolation and log frequency interpolation, the interpolation is computed as:
EMI Receiver Mode Reference
Meas Setup, Limits, Properties
:CALCulate:FSCan:LLINe[1]|2|...|6:TRACe 1|2|3
:CALCulate:FSCan:LLINe[1]|2|...|6:TRACe?
CALC:FSC:LLIN3:TRAC 2  Applies limit 3 to trace 2.
When the trace display is off, the trace is not tested. The trace is tested only when the trace display
is on and Test Limits (see
"Test Limits (All Limits)" on page
This matters when testing a trace or limit line for failure, via :CALC:FSC:LLIN3:FAIL? or
:CALC:FSC:TRAC2:FAIL?
1|1|2|2|3|3
Saved in instrument state.
1|2|3
A.07.00
Meas Setup, Limits, Properties
In square brackets, the state of Frequency Interpolation  then the state of Amplitude Interpolation,
separated by commas, as [Lin Frequency, Log Amplitude]
A.07.00
8  Frequency Scan Measurement
496) is on.
Meas Setup
485