Signal List - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

8  Frequency Scan Measurement
Recall
saved the limit line data to. Otherwise, the default recall directory is My Documents\EMC Limits and
Ampcor\Limits, which contains a set of preloaded limit files.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Range
Readback
Initial S/W Revision

Signal List

Selects the signal list as the data type to be imported. The next step is to select the Open key in the Import
Data menu. If there is signal in signal list, and you accessed this function from the front panel, the following
message will prompt you for confirmation to delete all signals from the signal list. Importing a Signal List
always replaces the data with the data that is already in the signal list.
When recalling a signal list, the current value of the Y-axis unit will be used to display the data. The limit to
delta values will be imported, you have to perform Update Δ limit in order to see the updated limit delta
based on current limit setting.
Key Path
Remote Command
Example
Notes
Status Bits/OPC
Dependencies
Initial S/W Revision
532
Recall, Export Data, Limit
:MMEMory:LOAD:LIMit:DDIRectory USER|PRELoaded
:MMEM:LOAD:LIM:DDIR USER Select user
:MMEM:LOAD:LIM:DDIR?
USER = User
PRELoaded = Preloaded
PRELoaded
Saved in instrument state.
User| Preloaded
"User"| "Preloaded"
A.13.00
Recall, Data
:MMEMory:LOAD:SLISt <filename>
MMEM:LOAD:SLIS "mySignalList2.csv"
Imports the Signal List from the file mySignalList2.csv in the default path.
If the file is empty, error –250.3005 is reported. If the file does not exist error –256 is reported. If
there is a mismatch between the file and the destination data type, an error is reported. –250.3003.
Then return to the Import Data menu and File Open dialog goes away.
Before loading a signal list file, any running scan sequence will be stopped unless it is in Scan Only.
Sequential - aborts the current measurement
A.07.00