Description - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

It is possible to change this setting after a limit line has been entered. When changing from On to Off or
vice-versa, the amplitude values in the limit line table change so that the limit line remains in the same
position for the current reference level settings of the instrument.
Key Path
Remote Command
Example
Couplings
Preset
State Saved
Range
Initial S/W Revision

Description

Provides a description of up to 60 characters by which you can easily identify the limit. The descriptions will
be stored in the exported file and can be displayed in the active function area by selecting them as the
active function, if desired to be in a screen dump.
Key Path
Remote Command
Example
Preset
State Saved
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision
Comment
Sets an ASCII comment field, which will be stored in an exported file. The comment can be displayed in the
active function area by selecting it as the active function, if desired to be in a screen dump. The Limits .csv
file supports this field.
Key Path
Remote Command
EMI Receiver Mode Reference
Meas Setup, Limits, Properties, Fixed/Relative
:CALCulate:LLINe[1]|2|...|6:AMPLitude:CMODe:RELative ON | OFF | 1 | 0
:CALCulate:LLINe[1]|2|...|6:AMPLitude:CMODe:RELative?
CALC:LLIN:AMPL:CMOD:REL ON  Makes limit line 1 relative to the reference level amplitude.
Pressing this button makes Reference level the active function.
Off, not affected by Mode Preset, preset by Restore Mode Defaults.
Saved in instrument state
On|Off
A.13.00
Meas Setup, Limits, Properties
:CALCulate:LLINe[1]|2|...|6:DESCription "string"
:CALCulate:LLINe[1]|2|...|6:DESCription?
CALC:LLIN:DESC "European Emissions"
"" (null String), not affected by Mode Preset, preset by Restore Mode Defaults.
Saved in instrument state
:CALCulate[1]|2:LIMit[1]|2|...|6:NAME
The suffix [1]|2 is irrelevant to the measurement, they are added solely for remote language
compatibility.
A.13.00
Meas Setup, Limits, Properties
:CALCulate:LLINe[1]|2|...|6:COMMent "string"
:CALCulate:LLINe[1]|2|...|6:COMMent ?
10  Monitor Spectrum Measurement
Meas Setup
795