Trig Delay - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

7  Trigger Functions
Trigger
Remote Command
Example
Preset
State Saved
Backwards Compatibility
Notes
Initial S/W Revision

Trig Delay

Controls a time delay during that the analyzer will wait to begin a sweep after meeting the trigger criteria.
You can use negative delay to pre-trigger the instrument in the time domain or FFT, but not in swept spans.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Default Unit
Backwards Compatibility
Notes
Initial S/W Revision
368
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
TRIG:SLOP NEG
POSitive
Saved in instrument state
In ESA/PSA, the Trigger Slope was global to all triggers.  In the X-Series, the slope can be set
individually for each Trigger Source.  For backward compatibility, the global SLOPe command
updates all instances of trigger slope (VID, LINE, EXT1, EXT2, TV, RFB). The query returns the trigger
slope setting of the currently selected trigger source.
Prior to A.02.00
Trigger, Video
:TRIGger[:SEQuence]:VIDeo:DELay <time>
:TRIGger[:SEQuence]:VIDeo:DELay?
:TRIGger[:SEQuence]:VIDeo:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:VIDeo:DELay:STATe?
TRIG:VID:DEL:STAT ON
TRIG:VID:DEL 100 ms
Video trigger delay may be set to negative values, in time domain, FFT and even swept. It makes
intuitive sense in time domain and works well in FFT mode where the bandwidth of the filter before
the video trigger is about 1.25 span. In swept spans, negative settings of Trig Delay are treated as a
zero setting within the internal hardware and the advisory message "Neg. Trig Delay unavailable in
Swept Mode, zero delay used." is generated when such a delay is set.
Off, 1 us
Saved in instrument state
–150 ms
+500 ms
s
! For backward compatibility with VSA/PSA comms apps
:TRIGger[:SEQuence]:IF:DELay
:TRIGger[:SEQuence]:DELay
The legacy :TRIGger[:SEQuence]:DELay command affects the delay for the VID, LINE, EXT1, EXT2,
and RFB triggers.
Prior to A.02.00
EMI Receiver Mode Reference

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents