Questionable Power Enable - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Mode
Remote Command
Example
Preset
Status Bits/OPC
dependencies
Initial S/W Revision

Questionable Power Enable

This command determines which bits in the Questionable Power Condition Register will set bits in the
Questionable Power Event register, which also sets the Power Summary bit (bit 3) in the Questionable
Register. The variable <integer> is the sum of the decimal values of the bits you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
Status Bits/OPC
dependencies
Initial S/W Revision
Questionable Power Event Query
This query returns the decimal value of the sum of the bits in the Questionable Power Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
Remote Command
Example
Preset
Status Bits/OPC
dependencies
Initial S/W Revision
EMI Receiver Mode Reference
All
:STATus:QUEStionable:POWer:CONDition?
STAT:QUES:POW:COND?
0
Sequential command
Prior to A.02.00
All
:STATus:QUEStionable:POWer:ENABle <integer>
:STATus:QUEStionable:POWer:ENABle?
STAT:QUES:POW:ENAB 32 50 MHz Input Pwr too High for Cal will be reported to the Power
Summary of the Status Questionable register.
32767
0
32767
Sequential command
Prior to A.02.00
All
:STATus:QUEStionable:POWer[:EVENt]?
STAT:QUES:POW?
0
Sequential command
Prior to A.02.00
3  Programming the Analyzer
STATus Subsystem