Clear Title - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Press Change Title to enter a new title through the alpha editor. Press Enter or Return to complete the
entry. Press ESC to cancel the entry and preserve your existing title.
The display title will replace the measurement name. It remains for this measurement until you press
Change Title again, or you recall a state, or a Preset is performed. A title can also be cleared by pressing
Title, Clear Title.
Notice the inclusion of the <measurement> parameter in the command below. Because each
measurement remembers the Display Title, the command must be qualified with the measurement name.
For the Swept SA measurement this is not the case; no <measurement> parameter is used when changing
the Display Title for the Swept SA measurement.
Key Path
Mode
Remote Command
Example
Notes
Preset
State Saved
Initial S/W Revision

Clear Title

Clears a title from the front-panel display. Once cleared, the title cannot be retrieved. After the title is
cleared, the current Measurement Name replaces it in the title bar.
Key Path
Example
Notes
Preset
Initial S/W Revision
EMI Receiver Mode Reference
View/Display, Display, Title
All
:DISPlay:<measurement>:ANNotation:TITLe:DATA <string>
:DISPlay:<measurement>:ANNotation:TITLe:DATA?
DISP:ANN:TITL:DATA "This Is My Title"
This example is for the Swept SA measurement in the Spectrum Analyzer mode. The SANalyzer
<measurement> name is not used.
DISP:ACP:ANN:TITL:DATA "This Is My Title"
This example is for Measurements other than Swept SA.
Both set the title to: This Is My Title
Pressing this key cancels any active function.
When a title is edited the previous title remains intact (it is not cleared) and the cursor goes at the
end so that characters can be added or BKSP can be used to go back over previous characters.
No title (measurement name instead)
Saved in instrument state.
Prior to A.02.00
View/Display, Display, Title
The following commands clear the title and restore the measurement's original title:
DISP:ANN:TITL:DATA ""
This example is for the Swept SA measurement in the Spectrum Analyzer mode. The SANalyzer
<measurement> name is not used.
DISP:ACP:ANN:TITL:DATA ""
This example is for ACP; in measurements other than Swept SA the measurement name is required.
Uses the :DISPlay:<measurement>:ANNotation:TITLe:DATA <string> command with an empty string
(in the Swept SA, the <measurement> is omitted).
Performed on Preset.
Prior to A.02.00
9  Strip Chart Measurement
View/Display
719