Limit Selection

These keys let you pick which Limit Line to import the data into; either 1, 2, 3, 4, 5 or 6. The default is the
currently selected limit.  Once selected, the key returns back to the Import Data menu and the selected
Limit Line number is annotated on the key. Now you have selected exactly what needs to be imported. In
order to trigger an import of the selected Limit Line, you must select the Open key in the Import Data menu.
An example of using this menu is: If you select 2 and continue to the File Open menu, Limit Line 2 will be
imported from the file selected or entered in File Name option in the File Open dialog.
Key Path
Notes
Preset
State Saved
Initial S/W Revision
Default Directory
Specifies the default directory used for loading of Limit. If user is selected, the default recall directory is the
default (My Documents\EMI\data\xxx\Limits where xxx is the measurement name) or the last directory you
saved the limit line data to. Otherwise, the default recall directory is My Documents\EMC Limits and
Ampcor\Limits, which contains a set of preloaded limit files.
Key Path
Remote Command
Example
Notes
834
myLimitLine2.csv in the current path. The default path is depending on the Limit Default Directory
selection.
Errors are reported if the file is empty or missing, or if the file type does not match, or if there is a
mismatch between the file type and the destination data type.  If any of these occur during manual
operation, the analyzer returns to the Import Data menu and the File Open dialog goes away.
This key will only appear if you have the proper option installed in your instrument.
When a limit line is loaded from mass storage, it is automatically turned on.  This allows you to see
it, thus confirming the load. The Margin settings will match those set when the limit was saved
1|2|3|4|5|6
Sequential - aborts the current measurement
A.02.00
Recall, Data, Limit Line
Auto return
Not part of Preset, but is reset to LLINE1 by Restore Mode Defaults; survives a shutdown.
The selected limit number is saved in instrument state
A.02.00
Recall, Export Data, Limit
:MMEMory:LOAD:LIMit:DDIRectory USER|PRELoaded
:MMEM:LOAD:LIM:DDIR USER Select user
:MMEM:LOAD:LIM:DDIR?
USER = User
PRELoaded = Preloaded