6  System Functions

System

not put into the SCPI error queue by default. Normally the status bit is the only way for detecting these
errors remotely.
It is possible to enable Input Overload reporting to the SCPI queue, by issuing the
:SYSTem:ERRor:OVERload ON command. To return to the default state, issue the
:SYSTem:ERRor:OVERload OFF command. In either case, Input Overloads always set the status bit.
For versions of firmware before A.10.01, the Input Overload was only a Warning and so was never available
in the SCPI queue, although it did set the status bit.  For A.10.01 and later, the Input Overload is an error
and can be enabled to the SCPI queue using this command.
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
System
The System screen is formatted into three groupings: product descriptive information, options tied to the
hardware, and software products:
The Previous Page is grayed-out if the first page of information is presently displayed. The Next Page menu
key is grayed-out if the last page is information is presently displayed.
Key Path
276
SCPI only
:SYSTem:ERRor:OVERload[:STATe] 0|1|OFF|ON
:SYST:ERR:OVER 1  Enable overload errors
Set to OFF by Restore Misc Defaults (no Overload errors go to SCPI)
Saved in instrument state.
A.10.01