Couple Markers - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Example
Notes
Couplings
Preset
State Saved
Range
Readback Text
Initial S/W Revision

Couple Markers

When this function is on, moving any marker causes an equal X Axis movement of every other marker that
is not off. By "equal X Axis movement" we mean that we preserve the difference between each marker's X
Axis value (in the fundamental x-axis units of the trace that marker is on) and the X Axis value of the marker
being moved (in the same fundamental x-axis units).
This may result in markers going off screen.
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
All Markers Off
Turn all markers off.
Key Path
Remote Command
Example
Initial S/W Revision
EMI Receiver Mode Reference
CALC:FSC:MARK:TRAC 1
CALC:FSC:MARK:TRAC?
A marker may be placed on a blanked and/or inactive trace, even though the trace is not visible
and/or updating.
Sending the remote command causes the addressed marker to become selected.
1
Saved in instrument state.
1|2|3
Trace N where N is the trace number to which the marker is currently assigned.
A.07.00
Marker
:CALCulate:FSCan:MARKer:COUPle[:STATe] OFF|ON|0|1
:CALCulate:FSCan:MARKer:COUPle[:STATe]?
CALC:FSC:MARK:COUP ON  Sets Couple Markers on.
Off, presets on Mode Preset and All Markers Off
Saved in instrument state.
A.07.00
Marker
:CALCulate:FSCan:MARKer:AOFF
CALC:FSC:MARK:AOFF
A.07.00
8  Frequency Scan Measurement
Marker