Calibrate - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

LFR1
Turns on low frequency rejection, rejecting signal component lower than 1.7 Hz.
Key Path
Example
Initial S/W Revision
LFR2
Turns on low frequency rejection, rejecting signal component lower than 0.14 Hz.
Key Path
Example
Initial S/W Revision

Calibrate

Invokes the guided probe calibration. The guided probe calibration is context sensitive and depends on the
channel (I or Q) and the Differential Input state. The calibration is only performed on the selected channel.
When Differential Input is on, both the probe attached to the main port and the probe attached to the
complementary port are calibrated. When Differential Input is off, only the probe attached to the main port
is calibrated. See
"I/Q Guided Calibration " on page
Key Path
Readback Text
Initial S/W Revision
Clear Calibration
Clears the calibration data for the current port and probe. It does not clear the data for other probe types
or other ports. If the sensed probe has EEPROM identification, only the data for that specific probe is
cleared. After this command has completed, the probe calibration state will be the same as if no probe
calibration had ever been performed for the specified channel and probe. The probe attenuation will be the
default value for that probe type and the Cable Calibration frequency response corrections will be used.
This command is dependent on the Differential Input state. When Differential Input is on, both the data for
EMI Receiver Mode Reference
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Turn on low frequency rejection on the I channel for frequencies lower than 1.7 Hz
INP:COUP:I LFR1
Prior to A.02.00
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Turn on low frequency rejection on the I channel for frequencies lower than 0.14 Hz
INP:COUP:I LFR2
Prior to A.02.00
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
The last calibration date, or if no calibration exists, "(empty)".
Last: <cal date>
<cal time>
Example:
Last: 8/22/2007
1:02:49 PM
Prior to A.02.00
208.
4  Input/Output Functions
Input/Output
175

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents