Max Hold

In Max Hold type the analyzer maintains and displays a max hold trace, which represents the maximum
data value on a point-by-point basis of the new trace data and previous trace data.
Key Path
Example
Dependencies
Couplings
Preset
State Saved
Initial S/W Revision
Min Hold
In Min Hold type the analyzer maintains and displays a min hold trace, which represents the minimum data
value on a point-point basis of the new trace data and previous trace data.
Key Path
Example
Dependencies
EMI Receiver Mode Reference
Trace/Detector
TRAC:FSC:TYPE AVER
This key is grayed out when a scan sequence is running except during single range scanning. If the
grayed out key is pressed, an advisory message is generated. If the equivalent remote command is
sent, this same message is generated as part of a "–221, Settings conflict" warning.
Affected by # of Scans (Avg Number).
Whenever you press Average or send the equivalent remote command,
Update is set to On and Display is set to On.
After a Preset, any trace that is in Trace Average is cleared (all trace points set to mintracevalue).
Saved in instrument state.
A.07.00
Trace/Detector
TRAC:FSC:TYPE MAX
This key is grayed out when a scan sequence is running except during single range scanning. If the
grayed out key is pressed, an advisory message is generated. If the equivalent remote command is
sent, this same message is generated as part of a "–221, Settings conflict" warning.
Affected by # of Scans (Avg Number).
Whenever you press Max Hold or send the equivalent remote command, Update is set to On and
Display is set to On.
After a Preset, any trace that is in Max Hold is cleared (all trace points set to
mintracevalue).
Saved in instrument state.
A.07.00
Trace/Detector
TRAC:FSC:TYPE MIN
This key is grayed out when a scan sequence is running except during single range scanning. If the
grayed out key is pressed, an advisory message is generated. If the equivalent remote command is
sent, this same message is generated as part of a "–221, Settings conflict" warning.