Min
Max
Default Unit
Initial S/W Revision
Y Offset
Offsets all segments in the limit line by some specified amplitude.
Key Path
Remote Command
Example
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
Apply Offset to Limit Table
Adds the X and Y offsets to each point in the limit table, then resets the X and Y offset values to zero. This
has no effect on the position of the limit trace.
For example, if the X offset is –10 MHz and the Y offset is 1 dB, the values in the limit table will be updated
as follows: 10 MHz will be subtracted from each X value, 1 dB will be added to each Y value. The offset
values will then be reset to zero. The limit trace will not be moved and the limit table will be updated to
accurately reflect the currently displayed limit trace. 
Key Path
Remote Command
Example
Initial S/W Revision

Scale X Axis

Matches the X Axis to the selected Limit, as close as possible.
EMI Receiver Mode Reference
–500 GHz
500 GHz
Determined by X axis scale.
A.07.00
Meas Setup, Limits, Edit, Offset
:CALCulate:FSCan:LLINe[1]|2|...|6:OFFSet:Y <rel_ampl>
:CALCulate:FSCan:LLINe[1]|2|...|6:OFFSet:Y?
CALC:FSC:LLIN:OFFS:Y –3 dB  Sets the Y axis offset to –3 dB.
CALC:FSC:LLIN:OFFS:UPD  Applies the Y axis offset to all points in the limit line, then reset the Y axis
offset to zero.
0 dB
Saved in instrument state.
–9.9E+37
9.9E+37
dB
A.07.00
Meas Setup, Limits, Edit, Offset
:CALCulate:FSCan:LLINe[1]|2|...|6:OFFSet:UPDate
CALC:FSC:LLIN3:OFFS:UPD  Updates the limit table 3 to reflect the X and Y offsets, then resets the
offsets to zero.
A.07.00
8  Frequency Scan Measurement
Meas Setup
493