I-Bar Port - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

4  Input/Output Functions
Input/Output
Key Path
Initial S/W Revision
Next
Perform the I port calibration.
Key Path
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired
calibration data.
Key Path
Notes
Initial S/W Revision

I-bar Port

The I-bar port calibration is performed with the probe body attached to the front panel's I-bar port and the
probe tip connected via an adapter to the Cal Out port. The I-bar probe calibration is only available for
passive probes with Differential On. The guided calibration will show a diagram of the required
connections.
Show Adapter
Show a connection diagram and instructions for the probe and adapter. See
page
220.
Key Path
Notes
216
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
:CALibration:IQ:PROBe:I
CAL:IQ:PROB:I
The I port must be connected to the Cal Out port before issuing the SCPI command.
The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not
reset by any preset or restore data commands.
No
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
Using the Exit button will not restore the calibration data to the state prior to entering the guided
calibration. Once a port is calibrated the data is stored immediately and the only way to change it is
to redo the calibration step.
When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see
"Exit Confirmation" on page 220
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
Either a passive or an active probe adapter diagram will be shown, depending on the type of probe
).
"Show Adapter Screen" on
EMI Receiver Mode Reference

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents