Questionable Calibration Register - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

3  Programming the Analyzer
STATus Subsystem
Remote Command
Example
Preset
Min
Max
Status Bits/OPC
dependencies
Initial S/W Revision

Questionable Calibration Register

"Questionable Calibration Condition " on page 116
"Questionable Calibration Enable " on page 116
"Questionable Calibration Event Query " on page 117
"Questionable Calibration Negative Transition " on page 117
"Questionable Calibration Positive Transition " on page 118
Questionable Calibration Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration Condition
register.
The data in this register is continuously updated and reflects the current conditions.
Mode
Remote Command
Example
Preset
Status Bits/OPC
dependencies
Initial S/W Revision
Questionable Calibration Enable
This command determines which bits in the Questionable Calibration Condition Register will set bits in the
Questionable Calibration Event register, which also sets the Calibration Summary bit (bit 8) in the
Questionable Register. The variable <integer> is the sum of the decimal values of the bits you want to
enable.
116
:STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
STAT:QUES:PTR 16
Temperature summary 'questionable asserted' will be reported to the Status Byte Register.
32767
0
32767
Sequential command
Prior to A.02.00
All
:STATus:QUEStionable:CALibration:CONDition?
STAT:QUES:CAL:COND?
0
Sequential command
Prior to A.02.00
EMI Receiver Mode Reference