Key Path
Remote Command
Example
Notes
Initial S/W Revision
Offset
Enters a menu that allows you to offset the limit trace by a specified frequency, time, or amplitude. The
offsets will be immediately applied to the limit trace for display and failure calculation. The offset can also
be applied to the points in the limit line.
Key Path
Initial S/W Revision
X Offset
Offsets the limit trace by some specified frequency (for Frequency-based limit lines) or a time (for time-
based limit lines).
Key Path
Remote Command
Example
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
Y Offset
Offsets all segments in the limit line by some specified amplitude.
Key Path
EMI Receiver Mode Reference
Meas Setup, Limits, Edit
:CALCulate:LLINe[1]|2|...|6:BUILd TRACE1|TRACE2|TRACE3
CALC:LLIN2:BUIL TRACE1 Builds limit line 2 based on the data in trace 1. This will overwrite the
data in the table editor.
Auto return to Edit menu.
A.13.00
Meas Setup, Limits, Edit
A.13.00
Meas Setup, Limits, Edit, Offset
:CALCulate:LLINe[1]|2|...|6:OFFSet:X <freq>
:CALCulate:LLINe[1]|2|...|6:OFFSet:X?
CALC:LLIN:OFFS:X –50MHZ Sets the X axis offset to –50 MHz.
CALC:LLIN:OFFS:UPD Applies the X axis offset to all points in the limit line, then reset the X axis
offset to zero.
0 Hz if Limit X-Axis Unit is Frequency
0 S if Limit X-Axis Unit is Time
Saved in instrument state
–500 GHz
500 GHz
Determined by X axis scale.
A.13.00
Meas Setup, Limits, Edit, Offset
10 Monitor Spectrum Measurement
Meas Setup