Rf Coupling - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Key Path
Remote Command
Example
Couplings
Preset
State Saved
Readback
Initial S/W Revision

RF Coupling

Specifies alternating current (AC) or direct current (DC) coupling at the analyzer RF input port. Selecting AC
coupling switches in a blocking capacitor that blocks any DC voltage present at the analyzer input. This
decreases the input frequency range of the analyzer, but prevents damage to the input circuitry of the
analyzer if there is a DC voltage present at the RF input.
In AC coupling mode, you can view signals below the corner frequency of the DC block, but below a certain
frequency the amplitude accuracy is not specified. The frequency below which specifications do not apply
is:
Lowest Freq for
meeting specs when
X-Series Model
AC coupled
N9000A–503/507
100 kHz
N9000A-C75 Input 2
1 MHz
N9000A–513/526
10 MHz
N9010A
10 MHz
N9020A
10 MHz
N9030A
10 MHz
Some amplitude specifications apply only when coupling is set to DC. Refer to the appropriate amplitude
specifications and characteristics for your analyzer.
EMI Receiver Mode Reference
Input/Output, RF Input
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
CORR:IMP 75 sets the input impedance correction to 75 ohms.
CORR:IMP?
In the N9000A option C75, when RF Input 2 is selected, the Input Z Correction will automatically
change to 75 ohms.  You may then change it to whatever is desired.  When the main RF Input is
selected, the Input Z Correction will automatically change to 50 ohms.  You may then change it to
whatever is desired. 
This is unaffected by a Preset but is set to 50 ohms on a "Restore Input/Output Defaults" or "Restore
System Defaults->All"
Some instruments/options may have 75 ohms available.
Saved in instrument state
50 Ω or 75 Ω. Current setting reads back to the RF key.
Prior to A.02.00
Lowest Freq for
meeting specs when
DC coupled
n/a
n/a
9 kHz
9 kHz
20 Hz
3 Hz